6515--NX EQ Endoscope Drying and Storage Cabinets - Vertical

SOL #: 36C10G25R0044Combined Synopsis/Solicitation

Overview

Buyer

Veterans Affairs
Veterans Affairs, Department Of
STRATEGIC ACQUISITION CENTER FREDERICKSBURG (36C10G)
FREDERICKSBURG, VA, 22408, United States

Place of Performance

Place of performance not available

NAICS

Surgical Appliance and Supplies Manufacturing (339113)

PSC

Medical And Surgical Instruments, Equipment, And Supplies (6515)

Set Aside

No set aside specified

Timeline

1
Posted
Jan 15, 2026
2
Submission Deadline
Sep 26, 2025, 3:59 AM

Qualification Details

Fit reasons
  • NAICS alignment with historical contract wins in similar service areas.
  • Scope strongly matches core technical capabilities and delivery model.
Risks
  • Past performance thresholds may require one additional teaming partner.
  • Potential clarification needed on staffing minimums before bid/no-bid.
Next steps

Validate eligibility requirements, assign capture owner, and schedule partner outreach to confirm teaming strategy before submission planning.

Amendment to Combined Synopsis-Solicitation

36C10G25R0042 0002

The purpose of this amendment is to:

Extend the Due Date for the solicitation.

The Due Date of the Combined Synopsis-Solicitation is extended to Thursday, 09/25/2025 at 11:59 PM ET.

Add responses to the below questions:

Question 1: Salient Characteristic 3 states The Endoscope Drying and Storage Cabinet shall be a stand-alone cabinet with the ability to expand scope capacity with an add on expansion unit. The verbiage add on expansion unit is only used by Steris. Other manufacturers allow for expansion through multiple units being set up together. Therefore, can you expand on what is required for an expansion unit? Can it be two units that are housed together in the same room?

CHANGE: Replace add on expansion unit with additional unit . The requirement will be the same as it is for the original unit. The additional unit would just expand the capacity to hold additional scopes.

Question 2: CLIN 0001 calls for a 6500 Ser HEPA Drying CBNT, 12 Scope Main Unit. Not all other manufacturers have 12 scope increments. Would another value, like 10 or 14, be acceptable?

CHANGE: The 12 Scope Main Unit is preferable; however, 10 and 14 Scope Main Units are tolerable (can serve as substitutes when the 12 is not met).

All other terms and conditions of the solicitation remain unchanged.

People

Points of Contact

Contract SpecialistDaleta ColesPRIMARY

Files

Files

Download
Download
Download
Download
Download
Download
Download
Download
View
Download
Download
Download
Download

Versions

Version 1Viewing
Combined Synopsis/Solicitation
Posted: Jan 15, 2026
6515--NX EQ Endoscope Drying and Storage Cabinets - Vertical | GovScope