Annual Preventive Maintenance and Certification for Biological Safety Cabinets (Class II, Type A2), Chemical Fume Hood, and PCR Cabinets/Clean Benches

SOL #: N6311626Q0516Sources Sought

Overview

Buyer

Dept Of Defense
Dept Of The Navy
NAVAL HEALTH RESEARCH CENTER
SAN DIEGO, CA, 92106, United States

Place of Performance

San Diego, CA

NAICS

Testing Laboratories and Services (541380)

PSC

Equipment And Materials Testing: Instruments And Laboratory Equipment (H266)

Set Aside

No set aside specified

Timeline

1
Posted
Feb 10, 2026
2
Response Deadline
Feb 17, 2026, 7:59 AM

Qualification Details

Fit reasons
  • NAICS alignment with historical contract wins in similar service areas.
  • Scope strongly matches core technical capabilities and delivery model.
Risks
  • Past performance thresholds may require one additional teaming partner.
  • Potential clarification needed on staffing minimums before bid/no-bid.
Next steps

Validate eligibility requirements, assign capture owner, and schedule partner outreach to confirm teaming strategy before submission planning.

Quick Summary

The Department of the Navy / Naval Health Research Center is conducting a Sources Sought to identify capable vendors for Annual Preventive Maintenance and Certification of laboratory equipment in San Diego, CA. This includes Biological Safety Cabinets, a Chemical Fume Hood, and PCR Cabinets/Clean Benches. Responses are due February 17, 2026.

Scope of Work

The requirement is for annual preventive maintenance (PM) and certification for a total of twenty-two (22) pieces of laboratory equipment located at the Naval Health Research Center. Services include:

  • Biological Safety Cabinets (Class II, Type A2) (Qty: 18): Testing and certification in accordance with NSF/ANSI 49 standards. Services must include downflow velocity profiling, inflow velocity measurement, HEPA filter integrity testing, smoke pattern testing, site installation assessment, and alarm function verification. Decontamination of the work area is also required.
  • Chemical Fume Hood (Qty: 1): Testing and certification in accordance with OSHA 1910.1450 and CAL OSHA Title 8 5154.1. Services must include face velocity measurement, smoke visualization testing, and alarm function verification.
  • PCR Cabinets / Clean Benches (Qty: 3): Testing and certification to ensure ISO 5 (Class 100) air quality conditions are met. Services must include HEPA filter integrity testing, non-viable particle counting, and UV light function verification.

Upon completion, the vendor must provide detailed Certification Reports for each piece of equipment, documenting tests performed, standards used, results, and any adjustments. Physical Certification/Testing Labels indicating the date of testing/certification and the next scheduled date must be affixed to each item.

Contract & Timeline

  • Type: Sources Sought / Market Research
  • PSC: H266 - Equipment And Materials Testing: Instruments And Laboratory Equipment
  • Set-Aside: None specified
  • Response Due: February 17, 2026, 07:59 AM PST
  • Published: February 10, 2026

Submission Instructions

Interested sources should submit capabilities information via email to David L. Hill (david.l.hill4.civ@health.mil). Submissions must reference N6311626Q0516 and include Company Name, Full Address, CAGE Code, and Point of Contact information (Name, Phone Number, E-Mail address).

People

Points of Contact

David L. HillPRIMARY

Files

Files

No files attached to this opportunity

Versions

Version 1Viewing
Sources Sought
Posted: Feb 10, 2026
Annual Preventive Maintenance and Certification for Biological Safety Cabinets (Class II, Type A2), Chemical Fume Hood, and PCR Cabinets/Clean Benches | GovScope