The Department of Veteran s Affairs, Altoona Medical Center, Altoona, PA 16602, has a requirement for Preventative Maintenance and Repair (PM&R) for Ameriwater brand SPD Reverse Osmosis equipment.
The North American Industry Classification System (NAICS) code being considered for this procurement is 811210, Electronic Precision and Equipment Repair and Maintenance, and the Product Service Code (PSC) is J064 Maintenance, Repair, and Rebuilding Equipment-Water Purification and Sewage Treatment Equipment. The Small Business Size Standard is $34 million. Responses to this notice should include the following:
Company name
SAM UEI
Company s address
Point of Contact Information (i.e., title, phone number, and email address)
Company s capability to meet this requirement and any pertinent information which demonstrates the company s ability to meet the above requirement.
GSA Federal Supply Schedule contract number or other applicable contract vehicle contract number, if within scope of this effort (if applicable).
Company s type of business (small or large)
Socio-economic status
If applicable, Service-Disabled Veteran Owned Small Businesses (SDVOSB) and Veteran Owned Small Business (VOSB) must be verified in www.VetBiz.gov. Responses should provide enough information to determine if the vendor can comply with the limitations on subcontracting as defined in VAAR 852.219-74 and all applicable CFR references where applicable.
All responses shall be in writing via email no later than 10:00 am (ET) November 14, 2025, to Emily Hall at emily.hall5@va.gov.
Facsimile or telephone responses will not be accepted. There will be no question-and-answer period for this sources sought.
Simply responding as an interested party does constitute your company as a source. Offeror must provide the above requested information. After reviewing the responses to this sources sought notice, a solicitation announcement may be published on the SAM or GSA website in the near future. Responses to this sources sought announcement are not considered adequate responses to the solicitation announcement. All interested offerors will have to respond to the forthcoming solicitation announcement, in addition to responding to this sources sought notice. All interested parties must be registered with System for Award Management (SAM) at https://www.sam.gov in order to be eligible for award of Government contracts.
Statement of Need
Background: The Altoona VA Medical Center wishes to procure a PM&R service contractor for SPD Reverse Osmosis equipment. All necessary labor must be performed by an authorized Ameriwater Service Provider capable of providing materials, equipment, parts and travel to perform inspections on the AMERIWATER ROS 800 GPD REVERSE OSMOSIS SYSTEM with Deionozer and Storage Tank (00HC-2050) at the James E. Van Xandt VA Medical Center, 2907 Pleasant Valley Boulevard, Altoona, PA 16602.
PERIOD OF PERFORMANCE: Base Plus 4 one-year option periods
Base Year: 12/03/2025-12/02/2026
Option Year 1: 12/03/2026-12/02/2027
Option Year 2: 12/03/2027-12/02/2028
Option Year 3: 12/03/2028-12/02/2029
Option Year 4: 12/03/2029 12/02/2030
QUARTERLY SERVICE:
-Inspect System for Leaks
-Verify Proper Functioning of RO Gauges &Instrumentation
-Monitor RO Membrane Performance
-Verify Proper Functioning of Storage Tank
System/ Level Controls
-Verify Proper Functioning of Distribution Pump System
-Verify Proper Functioning of Anti-Scalant Feed System
-Replace two Ameriwater Carbon Block Filter Cartridges 2.5" x 20"10 micron
-Check and Fill Antiscalant. Use customer stock to refill.
-RO Cleaning/Disinfection
SEMI-ANNUAL SERVICE TO INCLUDE:
-Minncare disinfection of the silex unit prior to replacement of (2) DI Sacks
-Disinfect the silex housings before
installing the new sacks
-Replace DI Packs (send out exhausted sacks to be regenerated for the next service. Includes replacement certificates)
*To be completed alongside a quarterly service*
ANNUAL SERVICE TO INCLUDE:
-Disinfection of Storage Tank
-Disinfection of 55 gallon storage tank
*To be completed alongside a quarterly service*
PARTS INCLUDED:
-AMERIWATER CHLOR PLUS CARBON
CARTRIDGES- CASE OF 6 (PN: 124009)
- AMERIWATER, PT-401, ANTI-SCALANT, 4
X 1 GALLON (PN: 153796)
- (TWO) AMERIWATER DI-2101S, STARTUP
STANDARD RECIRC. RESIN PACK, .5
CU.FT.
- (TWO) AMERIWATER DI-2101C,
REGENERATION CERTIFICATE,
STANDARD DI, .5 CU.FT.