Liquid Chromatography Mass Spectrometer

SOL #: W911S226U3135Combined Synopsis/Solicitation

Overview

Buyer

DEPT OF DEFENSE
Dept Of The Army
W6QM MICC-FT DRUM
FORT DRUM, NY, 13602-5220, United States

Place of Performance

Place of performance not available

NAICS

Analytical Laboratory Instrument Manufacturing (334516)

PSC

Chemical Analysis Instruments (6630)

Set Aside

Total Small Business Set-Aside (FAR 19.5) (SBA)

Timeline

1
Posted
Apr 30, 2026
2
Submission Deadline
May 15, 2026, 4:00 PM

Qualification Details

Fit reasons
  • NAICS alignment with historical contract wins in similar service areas.
  • Scope strongly matches core technical capabilities and delivery model.
Risks
  • Past performance thresholds may require one additional teaming partner.
  • Potential clarification needed on staffing minimums before bid/no-bid.
Next steps

Validate eligibility requirements, assign capture owner, and schedule partner outreach to confirm teaming strategy before submission planning.

Quick Summary

The Department of the Army, W6QM MICC-FT DRUM, has issued a Combined Synopsis/Solicitation for a Liquid Chromatography Mass Spectrometer. This requirement is a Total Small Business Set-Aside for the procurement of an integrated instrument for chemical analysis. The system must be delivered and installed at West Point, NY 10996. Responses are due by May 15, 2026, 4:00 PM ET.

Scope of Work

The solicitation seeks a combined liquid chromatography and mass spectrometer system. Key technical requirements include:

  • Integrated System: Both LC and MS components manufactured and serviced by the same company, controlled by a single unified software platform.
  • Analytical Capabilities: Must perform both qualitative and quantitative analysis in a single run, with software for data processing and acquisition. Features include automatic restart upon power failure and peak shape asymmetry detection.
  • Ionization & Detection: Requires a dual ion source (Electrospray Ionization (ESI) and Atmospheric Pressure Chemical Ionization (APCI)) and an off-axis, high-energy conversion dynode detector with a discrete multistage active film electron multiplier (without phosphorconversion).
  • Performance & Maintenance: Maximum nitrogen consumption of 15 L/min, autotuning for positive/negative ion modes, 15,000 u/sec scan speed (0.1u steps), 0.7u scan step resolution, and a mass range of 2-2000 m/z. The desolvation capillary must be removable and replaceable without breaking vacuum.
  • Logistics: Includes delivery and installation to West Point, NY, and a 1-year warranty on parts and labor.
  • Compliance: This is a Commercial Off-The-Shelf (COTS) item, exempt from DFARS 252.204-7012, and does not require CMMC 2.0 compliance.

Contract & Timeline

  • Opportunity Type: Combined Synopsis/Solicitation
  • Set-Aside: Total Small Business (FAR 19.5)
  • Product Service Code: 6630 - Chemical Analysis Instruments
  • Response Due Date: May 15, 2026, 4:00 PM ET
  • Published Date: April 30, 2026, 4:01 PM ET
  • Place of Performance: West Point, NY 10996

Contact Information

For inquiries, contact Carl Lundell at carl.lundell@westpoint.edu or 845.938.8298.