VPRO Max2 Low Temp Sterilizers

SOL #: 36C25526Q0154Sources Sought

Overview

Buyer

Veterans Affairs
Veterans Affairs, Department Of
255-NETWORK CONTRACT OFFICE 15 (36C255)
LEAVENWORTH, KS, 66048, United States

Place of Performance

Kansas City, MO

NAICS

Electromedical and Electrotherapeutic Apparatus Manufacturing (334510)

PSC

Medical And Surgical Instruments, Equipment, And Supplies (6515)

Set Aside

No set aside specified

Timeline

1
Posted
Jan 14, 2026
2
Response Deadline
Jan 21, 2026, 5:00 PM

Qualification Details

Fit reasons
  • NAICS alignment with historical contract wins in similar service areas.
  • Scope strongly matches core technical capabilities and delivery model.
Risks
  • Past performance thresholds may require one additional teaming partner.
  • Potential clarification needed on staffing minimums before bid/no-bid.
Next steps

Validate eligibility requirements, assign capture owner, and schedule partner outreach to confirm teaming strategy before submission planning.

2. SPECIFICATIONS
. Minimum of three low temperature sterilizers shall have a minimum individual dimensions of
Cabinet (DD): 33 x 40-15/16 x 72” (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38-
5/16 x 72” (838 x 973 x 1829 mm).
o Mounting single door, cabinet
o System installation requires no plumbing, ventilation or air
o supply – only a dedicated electrical connection is needed. A power cord is supplied for
this connection
o The facility space in which the vehicle can occupy is approx. 35’ x 110’.
. Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive
materials because temperatures within the load currently do not exceed 55ºC (131ºF)
and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can
also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized
hydrogen peroxide sterilizers, either with or without plasma, are currently available in the
United States.
. Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area
with a ventilation system capable of at least 10 air exchanges per hour.
. VPRO MaX2 must support utility requirements:
o Current 16 AMP
o Minimum Circuit Breaker 20 AMPS
o Use NEMA L15-20R Plug
o Provide grouped or ganged circuit protection and disconnect for phase 3 power as
required by NEC and local codes.
o Individual power shutoffs near each machine for servicing.